注:在鋰電池和有機金屬應用中,操作過程中揮發的酸/堿會腐蝕濕度傳感器,從而導致電阻變化。顯示的濕度值可能與實際濕度值不同。偏差程度取決于腐蝕性物質的濃度以及傳感器刷新過程的徹底性。在這種情況下,我們建議在保修期內將分析儀發送給我們進行免費校準。
Maintenance of the Moisture Analyzer
Attention: The phosphoric acid solution used for refreshing the functionality of the moisture sensor is corrosive. Please wear protective gloves and eyewear.
Disconnect the system cable before removing the moisture analyzer from the glove box
Unscrew the shroud (protective cage) cover from the moisture probe.
Apply distilled water on the sensor probe. Dry with a lint-free cloth (do not use paper towels). Handle the probe with care. Avoid excessive contact the surface of the probe.美國VTI手套箱微量水分析儀探頭MK-XTR-100使用注意事項
Moisten the sensor with the phosphoric acid solution by slowly dripping 10 to 15 drops directly on the probe while rotating the unit. Drip enough of the solution onto the probe until the surface is completely covered.
Allow the sensor to air-dry for four hours. Place a drop cloth under the probe to catch excess acid.
(©)Vacuum Technology Inc., 15 Great Republic Drive, Gloucester, MA, 01930 USA 2016
When completely dry, replace the protective shroud on the probe.
Confirm that the glove box is still within the pressure range of +5 mBar to +8 mBar before removing the blind flange cover and replacing the sensor in the port.
Reconnect the system cable after the moisture analyzer is reinstalled in the glove box.
On the touchscreen, touch “Analyzer” to activate the moisture analyzer.
Log the moisture data
Note: In lithium battery and in organometallic applications, volatilized acid/alkali during operation will corrode the moisture sensor which can cause a change in electrical resistance. The displayed moisture value might be different from the actual moisture value. The degree of deviation depends on the concentration of corrosive substances as well as the thoroughness of the sensor-refreshing process. In this case, we recommend sending the analyzer to us for a free calibration within the warranty period.